Appendix C
355
Temperature Characteristic Test Kit (Option 007)
Measuring temperature characteristic using sample program
C. T
e
m
per
atur
e
Char
acter
istic
Te
st Kit (
O
ption
007
)
1. Saving a measurement condition state file
This sample program allows you to perform measurement under up to 3 different state
conditions at each temperature point. At each temperature point, measurement is
performed while reproducing measurement conditions by loading the specified state file
(.sta). Prior to measurement, therefore, set measurement conditions and save the state file
after the execution of calibration/compensation.
NOTE
Keep the temperature of the temperature chamber constant within the range of 18
°
C to
28
°
C while creating a state file.
Step 1.
Set the measurement conditions of the instrument. For information on the setting
method/procedure, refer to Chapter 3, “Setting Measurement Conditions.”.
Step 2.
Execute calibration/compensation. For information on the execution procedure, refer to
“Calibration/compensation” on page 340.
Step 3.
Save the state file (.sta) into the following folder.
D:\TCTEST\STATE
If you want to perform measurement under other conditions, repeat steps 1 through 3.
2. Starting the program
Execute the sample program (VBA). For more information on the procedure, refer to
Chapter 13, “Use of Macros.” in the Programming Guide.
Step 1.
Load the sample program, D:\Tctest\tctest.lcr.
Step 2.
Execute the macro, Tctest.Start.
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...