Notice that <Area1> is the sensor name of an Area Sensor Type in the current inspection.
>> get blemish_history <Blemish1> minedgelength\x0D\x0A
<< OK\x0D\x0A
<< 22\x0D\x0A
Notice that <Blemish1> is the sensor name of a Blemish Sensor Type in the current inspection.
8.4.6 Command Channel Command Status Register
The command status is a verification of the command type.
Command Status
Value (16-bit integer)
Unknown
0
Read
1
Write
2
Execute
3
8.4.7 Command Channel Error Codes
Table 13: Plus TG Command Channel Error Codes
Numeric ID
Text ID
Description
00000
SUCCESS
Command processed successfully
10000
EMPTY_FRAME_RECEIVED
Indicates that the request was empty. The command channel
requires a command, any arguments, and an end-of-frame
delimiter.
10001
COMMAND_NOT_RECOGNIZED
The command specified is not recognized
10100
GROUP_MISSING
A Group ID must be specified immediately after the command
10101
GROUP_NOT_FOUND
The specified Group ID is invalid / unknown
10102
GROUP_ITEM_MISSING
A Group Item ID must be specified immediately after the Group
ID
10103
GROUP_ITEM_NOT_FOUND
The specified Group Item ID is invalid / unknown
10152
NOT_READABLE
Attempt to get a value that is not readable
10153
NOT_WRITEABLE
Attempt to set a value that is not writeable
10250
NOT_A_METHOD
Method ID specified is not a method
10251
WRONG_ARGUMENT_COUNT
Total method arguments specified do not match method
10252
COMMAND_NOT_FINISHED
Attempt to issue command when a previous command has not
finished
10300
INVALID_ARGUMENT_TYPE
Item ID specified must be a item (not a group or method)
10301
DATA_VALUE_MISSING
Command missing item's data value
10350
ARGUMENTS_DETECTED
Get command received with unneeded arguments
10351
INVALID_ARGUMENT_TYPE
Item ID specified must be a item (not a group or method)
10340
MINIMUM_VALUE_EXCEEDED
New item value is below the minimum
10341
MAXIMUM_VALUE_EXCEEDED
New items value is above the maximum
10500
DATA_SET_EMPTY
Data export operation returned no results.
10900
SENSOR_NOT_READY
Command specified requires sensor to be in the READY state.
10920
SENSOR_TYPE_NOT_ACTIVE
Command specified belongs to a different sensor type.
15000
VALUE_INVALID
Text value is invalid / unknown
15050
VALUE_INVALID
Text value is invalid - expecting True or False
15100
STRING_TOO_LONG
String value specified exceeds maximum allowable length
iVu Plus TG Gen2 Image Sensor
98
www.bannerengineering.com - Tel: 763.544.3164