116
8-3. DETAILS OF CONTROL COMMANDS
This section describes various control commands and parameters following after the commands.
<CR> and <CR><LF> denote termination code.
8-3-1. Initial setting
(1) [RES]
Content:
Initialize the setting value.
Format:
RESp<CR>
Parameter:
p
0: Input setting
1: Measuring condition setting
2: Waveform display setting
3: Initial setting
4: Optional setting
5: All settings
Return value format:
ok<CR><LF>
Description and others
This command is available even if the power is turned OFF.
(2) [RS-]
Content:
RS line test
Format:
RS-<CR>
Parameter:
None
Return value format:
ok<CR><LF>
Description and others
RS-232C<CR><LF> (For the RS-232C) or
RS-485<CR><LF> (For the RS-485)
(when the data is received correctly)
(3) [SYS]
Content:
Output model, etc.
Format:
SYS<CR>
Parameter:
None
Return value format:
WGA-910A-aa, xx. xx, yyyyyyyyy <CR><LF>
Serial No. (9-digit)
Character string of 0 to 9, A, B, C
Program version
If the indicated value has no 10th
digit, it is filled with ‘0’.
Optional model
The model
Description and others
Contents of the return value is as follows.
00: Without option.
01: With the BCD output.
02: With the D/A output.
03: With the RS-485.
04: With the CC-Link.
12: With the BCD and D/A output.