123
(21) [VRC]
Content:
To execute numeric value registering calibrations
Format:
VRC±x.xxxx<CR>
Parameter:
±x.xxxx
Initial value of transducers: -3.2000 to 3.2000 (mV/V)
Return value format:
ok<CR><LF>
E-00<CR><LF>(When an error occurs)
Description and others
The results of the command will be saved.
Reading command:
PRR011
Return value
x.xxxx<CR><LF>
Initial value of transducers:
(22) [EID]
Content:
Conduct TEDS-based automatic calibration.
Format:
EID<CR>
Parameter:
None
Return value format:
ok<CR><LF>
(when in normal)
E-13<CR><LF> (When an error occurs)
Description and others
Result of this command is overwritten.
Sets the excitation voltage and sensitivity registering calibration with the rated capacity
value, rated output value, recommended excitation voltage value, and unit of the TEDS
information.
Reading command
PRR010
Return value
±x.xxxx, ±yyyyyy, zz<CR><LF>
Unit
Rated display value
Rated output value
(23) [SSM]
Content:
Set the smoothing function.
Format:
SSMaa, b, c<CR>
Parameter:
aa, b, c
Analog filter
0: 1 Hz
1: 30 Hz
2: 300 Hz
3: None (1 kHz or more)
Minimum scale
0: 1
1: 2
2: 5
3: 10
4: 20
5: 50
6: 100
Moving average
00: None
01: 2 times
02: 4 times
03: 8 times
04: 16 times
05: 32 times
06: 64 times
07: 128 times
08 256 times
09: 512times
10: 1024 times 11: 2048 times
Return value format:
ok<CR><LF>
Description and others
Corresponding to the [/M] command.
Reading command
PRR020
Return value
aa, b, c<CR><LF>
Analog filter
Minimum scale
Moving average