129
(35) [OP1]
Content:
Set communication (RS-232C and RS-485)
Format:
OP1, a <CR>
Parameter:
OP1, a
Data Rate
0: 2400 bps
1: 4800 bps
2: 9600 bps
3: 19200 bps
Return value format:
ok<CR><LF>
Description and others
Corresponding to the [/M] command.
Reading command
PRR100
Return value
030, a <CR><LF>
Data Rate
(36) [OP2]
Content:
Set the BCD output
Format:
OP2, a, b, c, d, e, f<CR>
Parameter:
OP2, a, b, c, d, e, f
BCD classification (*2)
0: Peak value
1: Bottom value
Data form
0: BCD
1: Binary
Transmission speed
0: Approx. 16 times/sec.
1: Approx. 32 times/sec.
2: Approx. 64 times/sec.
3: Approx. 125 times/sec.
4: Approx. 250 times/sec.
5: Approx. 500 times/sec.
6: Approx. 1000 times/sec.
EOC logic
0: Negative logic
1: Positive logic
Polarity logic
0: Negative logic
1: Positive logic
Data logic
0: Negative logic
1: Positive logic
Return value format:
ok<CR><LF>
Description and others
Corresponding to the [/M] command.
Reading command
PRR100
Return value
a, b, c, d, e, f
BCD classification (*2)
Data form
Transmission speed
EOC logic
Polarity logic
Data logic
*2 For the [Block Peak-Bottom] and [Time Peak-Bottom] only. Or, you will not get the BCD or D/A classification
value.