Remote Operation
5-33
5.
SOURce: CURRent
Description:
It sets the output current (ampere).
Syntax:
SOURce:CURRent <NRf+>
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:CURR 1
It sets the output current to 1 amps.
SOUR:CURR 60.00
It sets the output current to 60.00 amps.
Query Syntax:
SOUR:CURR?
Return Parameter: <NRf+> [Unit Amp]
Query Example: SOUR:CURR?
It returns the current.
Return Example: 1.00
6.
SOURce:CURRent:LIMit:{HIGH/LOW}
Description:
It sets the output current range.
Syntax:
SOURce:CURRent:LIMIT:HIGH <NRf+>
SOURce:CURRent:LIMIT:LOW <NRf+>
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:CURR:LIMIT:HIGH 60.0 It sets the output current range
to 60A maximum.
SOUR:CURR:LIMIT:LOW 20.0 It set the low limit to 20A for
current output protection.
Query Syntax:
SOUR:CURR:LIMIT:HIGH?
SOUR:CURR:LIMIT:LOW?
Return Parameter: <NRf+> [Unit Amp]
Query Example: SOUR:CURR:LIMIT:HIGH? It returns the maximum range set
for current.
Return Example: 6.01
7.
SOURce:CURRent:PROTect:HIGH
Description:
It sets the current range for over current protection.
Syntax:
SOURce:CURRent:PROTect:HIGH <NRf+>
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:CURR:PROT:HIGH 60.0 It sets the high limit to 60A
for current output protection.
Query Syntax:
SOUR:CURR:PROT:HIGH?
Return Parameter: <NRf+> [Unit Amp]
Query Example: SOUR:CURR:PROT:HIGH?
It returns the high limit of
current protection.
Return Example: 6.01
8.
SOURce:CURRent:SLEW
Description:
It sets the rising or falling slew rate (amp/ms) of output current.
Syntax:
SOURce:CURRent:SLEW <NRf+>
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:CURR:SLEW 0.01 It sets the output current slew rate to
0.01 Amp/mS.
SOUR:CURR:SLEW 1.00 It sets the output current slew rate to
1.00 Amp/mS.
Query Syntax:
SOUR:CURR:SLEW?
Return Parameter: <NR f+> [Unit Amp/ms]
Query Example: SOUR:CURR:SLEW? It returns the current slew rate.
Return Example: 1.00