Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
5-34
9.
SOURce:CURRent:SLEWINF
Description:
It sets the I Slewrate to INF.
Syntax:
SOURce:CURRent:SLEWINF ENABLE
SOURce:CURRent:SLEWINF DISABLE
Parameter:
ENABLE/DISABLE
Example:
SOUR:CURR:SLEWINF ENABLE
It sets the I Slewrate to INF.
SOUR:CURR:SLEWINF DISABLE
It resets the I Slewrate and
returns to 1A/ms.
Query Syntax:
SOUR:CURR:SLEW?
Return Parameter: INF. Or <NRf+>[Unit = Ampere]
Query Example:
SOUR:CURR:SLEW?
It returns the setting.
Return Example: INF.
10. SOURce:POWer:PROTect:HIGH
Description:
It sets the over power point (Watt) of output power.
Syntax:
SOURce:POWer:PROTect:HIGH <NRf+>
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOURce:POWer:PROTect:HIGH 1260
It sets the over power
point 1260.
Query Syntax:
SOURce:POWer:PROTect:HIGH?
Return Parameter: <NR f+> [Watt]
Query Example: SOURce:POWer:PROTect:HIGH?
It returns the over
power setting value.
Return Example: 1.03
11. SOURce:DCON:{RISE/FALL}
Description:
It sets the DC_ON signal active point.
Syntax:
SOURce:DCON:RISE <NRf+>
SOURce:DCON:FALL <NRf+>
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:DCON:RISE 79.5
It sets the DC_ON rise to 79.5V.
SOUR:DCON:FALL 0.5
It sets the DC_ON fall to 0.5V.
Query Syntax:
SOUR:DCON:RISE?
SOUR:DCON:FALL?
Return Parameter: <NRf+> [Unit = Volt]
Query Example:
SOUR:DCON:RISE?
It returns the setting.
Return Example: 7.01
Note
The output must be OFF for setting.