Cryo-con Model 24C
Remote Programming Guide
LOOP
{1 | 2 | 3 | 4}
:AUTotune:STATus?
Queries the status of the autotune process. Return values are:
IDLE
- Autotune has not started.
RUNNING
-Autotune is running.
COMPLETE
-Autotune successfully completed.
FAILED
-Unable to generate PID values.
ABORT
-Aborted by operator intervention.
OVERTEMP commands
These commands are associated with the heater’s Over Temperature Disconnect
(OTD) feature. This is used to disconnect the heater if a specified temperature is
exceeded on any selected input channel.
OVERtemp:ENABle {ON | OFF}
Sets and queries the Over Temperature Disconnect enable. The OTD does
not function if disabled.
OVERtemp:SOURce {A | B | C | D}
Sets and queries the input channel that is used as the source for the Over
Temperature Disconnect feature.
OVERtemp:TEMPerature <temp>
Sets and queries the temperature used by the over temperature disconnect
feature. Note that this temperature has the same units of the source input
channel.
130