Cryo-con Model 24C
Appendix A: Installed Sensor Curves
The SENSORIX remote commands are used to query and edit sensors installed in
the controller. For example, the command:
INPUT B SENSORIX 34 would set input B to use the R400 sensor.
INPUT A: SENSORIX 1 would set input A to use the S900 diode.
INPUT A: SENSORIX 0 would turn input A off by setting the sensor to ‘none’.
SENSORIX 1:NAME? Returns the name string at index 1.
Factory installed sensors may not be edited by using these commands.
User Installed Sensor Curves
The user may install up to four custom sensors. This table shows the sensor index
and default name of the user curves:
User Curve
Sensor IX
Default Name
0
61
User Sensor 1
1
62
User Sensor 2
2
63
User Sensor 3
3
64
User Sensor 4
4
65
User Sensor 5
5
66
User Sensor 6
6
67
User Sensor 7
7
68
User Sensor 8
Using the above table, the SENSORIX commands can be used to address the user
curves. For example:
INPUT B SENSorix 62 assigns input B to user sensor #2.
SENSorix 64:NAMe? Returns the name string of user sensor 4
SENSorix 63:TYPe ACR sets the type of user sensor #3 to ACR.
i
NOTE:
Factory installed sensors are indexed from 0 to 60. User
installed sensors have index values from 61 to 68 corresponding to
user curves 1 through 8.
142