Remote Operation
5-41
SPECification:VOLTage?
Type: Channel-specific
Description:
Requests GO-NG result reference to voltage specification.
Query Syntax:
SPECification:VOLTage?
Query Example:
SPEC:VOLT?
Returns voltage GO-NG result to CC
and CR modes.
Return Parameters: <NR1>
Return Example:
0 (NG), 1 (GO)
SPECification:CURRent?
Type: Channel-specific
Description:
Requests GO-NG result reference to current specification.
Query Syntax:
SPECification:CURRent?
Query Example:
SPEC:CURR?
Returns current GO-NG result to CC
mode.
Return Parameters: <NR1>
Return Example:
0 (NG), 1 (GO)
SPECification?
Type: All
Channel
Description:
Requests GO-NG result reference to all channel specification.
Query Syntax:
SPECification?
Query Example:
SPEC?
Returns all channel GO-NG result.
Return Parameters: <NR1>
Return Example:
0 (NG), 1 (GO)
SPECification:VOLTage
Type: Channel-specific
Description:
Sets the voltage specification.
Syntax: SPECification:VOLTage:H
SPECification:VOLTage:L
SPECification:VOLTage:C
Parameters:
For valid value range, refer to respective specification.
Example:
SPEC:VOLT:H <NRf+> [suffix]
SPEC:VOLT:L <NRf+> [suffix]
SPEC:VOLT:C <NRf+> [suffix]