Sorensen MML Series Operation and Programming Manual
5-14
CHANnel:[LOAD]
Type: Channel
Specific
Description:
Selects a specific channel by which the coming channel-specific
command will be received and executed.
Syntax: CHANnel
<NRf+>
Parameters:
1 ~ 8
Example:
CHAN 1
Sets specific channel as "1".
CHAN MAX
Sets specific channel as "8".
CHAN MIN
Sets specific channel as "1".
Query Syntax:
CHAN?
CHAN? MAX
CHAN? MIN
Return Parameters: <NR1>
Query Example:
CHAN?
Returns current specific channel.
Return Example:
1
CHANnel:ACTive
Type: Channel
Specific
Description:
Enables or disables the load module.
Syntax:
CHANnel : ACTive ON. Enable the load module. The front panel
displays the measurement of voltage
and current.
CHANnel : ACTive OFF. Disable the load module. LCD on the
front panel displays OFF.
Parameter: ON/1,
OFF/0
Example:
CHAN : ACT ON
CHANnel:SYNCon
Type: Channel
Specific
Description:
Sets the load module to receive synchronized command action of
RUN ABORT or not.
Syntax:
CHANnel : SYNCon ON
CHANnel : SYNCon OFF
Parameters: ON/1,
OFF/0
Example:
CHAN : SYNC ON.
Sets the load module to receive
synchronized command action.
CHAN : SYNC OFF.
Sets the load module not to receive
synchronized command action.
Query Syntax:
CHAN : SYNC?
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com