Sorensen MML Series Operation and Programming Manual
5-38
Example:
RES:L1 20 OHM
Sets constant resistance = 20 ohm for
Load L1.
RES:L2 10 OHM
Sets constant resistance = 10 ohm for
Load L2.
RES:L1 MAX
Sets constant resistance = maximum L1
value for Load L1.
RES:L2 MIN
Sets constant resistance = minimum L2
value for Load L2.
Query Syntax:
RESistance:L1?
RESistance:L2?
RESistance:L1?
MAX
RESistance:L2?
MIN
Return Parameters: <NR2> [Unit=OHM]
Query Example:
RES:L1?
Returns the set resistance of the value
of Load L1.
Return Example:
10
RESistance:RISE/FALL
Type: Channel-specific
Description:
Sets resistive slew rate of constant resistance.
Syntax: RESistance:RISE
<NRf+>
[suffix]
RESistance:FALL
<NRf+>
[suffix]
Parameters:
For valid value range, refer to respective specification.
Example:
RES:RISE 2.5
Sets CR rise slew rate as 2.5A/
µ
S.
RES:FALL
1A/
µ
S
Sets CR fall slew rate as 1A/
µ
S.
RES:RISE MAX
Sets CR rise slew rate as the maximum
programmable value.
RES:FALL MIN
Sets CR fall slew rate as the minimum
programmable value.
Query Syntax:
RESistance:RISE?
RESistance:FALL?
RESistance:RISE?
MAX
RESistance:FALL?
MIN
Return Parameters: <NR2> [Unit=OHM]
Query Example:
RES:RISE?
Returns CR rise slew rate.
Return Example:
2.5