Command reference
R&S
®
ZNB/ZNBT
1204
User Manual 1173.9163.02 ─ 62
Example:
MMEM:LOAD:CKIT 'C:\Users\Public
\Documents\Rohde-Schwarz
\Vna\Calibration\Kits\New Kit 1
(123456).calkit'
Load the previously created cal kit file
New Kit 1 (123456).calkit
from the default cal kit direc-
tory.
CORR:CKIT:N50:LSEL 'New Kit 1', '123456'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in
New Kit 1 (123456).calkit
reads
New Kit 1
and that its label reads
123456
.
Manual operation:
See
"Add / Copy / Delete / Standards..."
[SENSe:]CORRection:CKIT:STANdard:LCATalog?
<CalKitName>, <KitLabel>
Returns a list of all standards in a given calibration kit.
Query parameters:
<CalKitName>
to query cal kit names and labels.
<KitLabel>
Label of the cal kit.
Example:
See
[SENSe<Ch>:]CORRection:CONNection:CATalog?
Usage:
Query only
[SENSe:]CORRection:CKIT:<StandardType>:WLABel
<ConnType>, <CalKitName>,
<CalKitLabel>, <StandardLabel>, <MinFreq>, <MaxFreq>, <DelayParam>,
<Loss>, <Z0>, <C0>, <C1>, <C2>, <C3>, <L0>, <L1>, <L2>, <L3>, OPEN |
SHORt | MATCh | <Resistance>[, <Port1>[, <Port2>]]
[SENSe:]CORRection:CKIT:<StandardType>:WLABel?
<ConnType>[,
<CalKitName>[, <CalKitLabel>[, <Port1>[, <Port2>]]]]
Defines the parameters of a non-ideal 1 port or 2-port calibration standard
<StandardType>
, where a particular cal kit can be addressed by name
and label
.
Apart from the additional
<CalKitLabel>
parameter, the syntax and semantics of this
command is identical to
Standard types and their parameters
.
Example:
CORR:CKIT:FOP:WLAB 'N 50 Ohm','New Kit
1','123456', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Define the properties of the Open (f) standard for a N 50 Ω con-
nector type in the calibration kit "New Kit 1" labelled "123456".
CORR:CKIT:FOP:WLAB? 'N 50 Ohm', 'New Kit 1',
'123456'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in the calibration kit.
Manual operation:
See
"Add / Copy... / Delete / View / Modify..."
SCPI command reference