Command reference
R&S
®
ZNB/ZNBT
1368
User Manual 1173.9163.02 ─ 62
SOURce:POWer:CORRection:PSELect
<CalPower>
Defines how to define the source power the R&S
ZNB/ZNBT uses to perform the first
calibration sweep of the source power calibration ("Reference Receiver Cal Power").
Parameters:
<CalPower>
CPOWer | PPOWer
CPOWer
Calculated port power result
PPOWer
Predefined port power value, see
*RST:
CPOWer
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
"Reference Receiver Cal Power"
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
<Boolean>
Enables or disables the source power calibration for channel
<Ch>
and for port number
<PhyPt>
. The setting command is disabled unless a source power calibration for the
analyzer port has been performed (
SOURce<Ch>:POWer:CORRection[:ACQuire]
).
The query always returns a result.
To enable or disable a source power calibration for an external generator use
SOURce<Ch>:POWer:CORRection:GENerator<Gen>[:STATe]
.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the source power calibration for
port number
<PhyPt>
.
*RST:
OFF
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
7.3.15.6
SOURce:POWer... (other)
The
SOURce:POWer...
commands define the power of the internal signal source.
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>:LEVel:OFFSet
SOURce<Ch>:POWer<PhyPt>:CORRection:LEVel:OFFSet
............................................. 1370
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:OFFSet
............................................... 1370
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:PERManent[:STATe]
............................. 1371
.............................................................................1372
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:STATe
................................................. 1372