EDH0162En1040 – 06/99
3.184
MM4005
YB — Negate variable
Usage
IMM
PGM
MIP
Syntax
xxYB
Parameters
Description
xx
[int]
— Variable number.
Range
xx
—
1
to
100
(integers) and
101
to
120
(floats)
.
Units
xx
— None.
Defaults
xx
Missing: Error O.
Out of range: Error O.
Floating point: Error A.
Description
This command negates the value of a variable. After executing this com-
mand, the value of variable
xx
takes the opposite sign.
Returns
None.
Errors
A
—
Unknown message code.
O
—
Variable number out of range.
Rel. Commands
TY
—
Read a variable.
YS
—
Initialize variable.
Example
3XX |
Clear program #3 from memory, if any.
3EP |
Activate program mode and enter following commands as program
#3.
7YS3 |
Initialize variable #7 to 3.
7YA2 |
Add 2 to variable #7; the new value for the variable is 5.
7YB
|
Negate variable #7; the new value for the variable is -5.
… |
… |
… |
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com