Network Operation and Remote Control
R&S
®
SMA100B
255
User Manual 1178.3834.02 ─ 03
Protocol
Remote control program
Socket
rssmw_init ("TCPIP:SMA100B-102030::5025::SOCKET", ...)
rssmw_SetFrequency (..., 2e9)
GPIB
rssmw_init ("GPIB::28::INSTR", ...)
rssmw_SetFrequency (..., 2e9)
11.4 Remote Access Settings
Network
Risk of network failure
Consult your network administrator before performing the following tasks:
●
Connecting the instrument to the network
●
Configuring the network
●
Changing IP addresses
●
Exchanging hardware
Errors can affect the entire network.
Default gateway
This address identifies the router on the same network as the instrument that is used to
forward traffic to destinations beyond the local network.
Instrument emulations
You can also remotely control the R&S
SMA100B via the command set of another sig-
nal generator, as, for example, of an HP generator. With this function you can, for
example, replace a signal generator with an R&S
SMA100B in an automated test
setup, without adjusting the command scripts used.
The R&S
SMA100B also covers command sets of Rohde
&
Schwarz signal genera-
tors, e.g. the R&S SMA100A, the R&S SMF100A, etc. To achieve optimal compatibility
when replacing an instrument, we recommend that you select the emulation command
set for the corresponding generator.
The selected instrument also defines the identification string that is retrieved with query
. If necessary, use the parameter
As any other parameter, the remote control command set can also be changed
remotely by the command
Remote Access Settings