Remote Control Commands
R&S
®
SMA100B
362
User Manual 1178.3834.02 ─ 03
Example:
SENS:LOGG:STAT ON
Activates recording of the power measurement of the first sen-
sor.
Manual operation:
See
:SENSe<ch>[:POWer]:OFFSet
<Offset>
Sets a level offset which is added to the measured level value after activation with
command
:SENSe<ch>[:POWer]:OFFSet:STATe
. The level offset allows, e.g. to
consider an attenuator in the signal path.
Parameters:
<Offset>
float
Range:
-100.0 to 100.0
*RST:
0
Default unit: dB
Example:
SENS1:POW:OFFS 10.0
Sets a level offset of 10 dB
Manual operation:
See
" Level Offset State,Level Offset"
:SENSe<ch>[:POWer]:OFFSet:STATe
<State>
Activates the addition of the level offset to the measured value. The level offset value is
set with command
.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SENS1:POW:OFFS 0.4dB
Sets a level offset of 0.4 dB
SENS1:POW:OFFS:STAT ON
A level offset of 0.4 dB is added to the measured value.
Manual operation:
See
" Level Offset State,Level Offset"
:SENSe<ch>[:POWer]:SNUMber?
Queries the serial number of the sensor.
Return values:
<SNumber>
string
Example:
SENS1:SNUM?
Queries the serial number.
Usage:
Query only
Manual operation:
See
"Sensor type and serial number"
SENSe, READ, INITiate and SLISt Subsystems