Remote Control Commands
R&S
®
SMA100B
407
User Manual 1178.3834.02 ─ 03
Return values:
<LfSinePeriod>
float
Range:
1E-6 to 100
Increment: 10E-9
*RST:
0.001
Default unit: s
Example:
See
Example "Configuring the LF generator"
Usage:
Query only
[:SOURce<hw>]:LFOutput:FREQuency:MANual
<Manual>
Sets the frequency of the subsequent sweep step if
Use a separate command for each sweep step.
Parameters:
<Manual>
float
You can select any value within the setting range, where:
STARt
[:SOURce<hw>]:LFOutput:FREQuency:
STOP
[:SOURce<hw>]:LFOutput:FREQuency:
Range:
STARt to STOP
Increment: 0.1
*RST:
1000
Example:
See
Manual operation:
See
[:SOURce<hw>]:LFOutput:FREQuency:MODE
<Mode>
Sets the mode for the output of the LF generator frequency, and determines the com-
mands to be used for frequency settings.
Parameters:
<Mode>
CW | FIXed | SWEep
CW|FIXed
Sets the fixed-frequency mode.
CW
and
FIXed
are synonyms.
To set the output frequency, use command
SWEep
Sets sweep mode.
To set the frequency, use the commands:
[:SOURce<hw>]:LFOutput:FREQuency:STARt
SOURce<hw>]:LFOutput:FREQuency:STOP
Or
[:SOURce<hw>]:LFOutput:FREQuency:MANual
*RST:
CW
SOURce Subsystem