Remote Control Commands
R&S
®
SMA100B
333
User Manual 1178.3834.02 ─ 03
12.8 DIAGnostic Subsystem
The
DIAGnostic
subsystem contains the commands used for instrument diagnosis
and servicing. SCPI does not define any
DIAGnostic
commands; the commands lis-
ted here are all device-specific. All
DIAGnostic
commands are query commands
which are not affected by
*RST
.
The test functions are intended for services purposes.
They are thus password-protected functions. Unlock the corresponding protection level
to access them, see
For more information, see R&S
SMA100B Service Manual.
Common suffixes
The following common suffixes are used in remote commands:
Suffix
Value range
Description
DIAGnostic<hw>
[1]
Optional suffix
Example: Programming example
The example lists the commands required to query assemblies and test points for diag-
nosis purposes.
// Query the modules available in the instrument
// and variant and revision state of a certain assembly
DIAGnostic1:BGINfo:CATalog?
// FRONT_AF,PSU300,IFB.BV,MB2_AF,...
DIAGnostic1:BGINfo? "PSU300"
// PSU300 2118.2067.02 01.00 100000
// Query the test points available in the instrument
// and trigger the measurement in a selected test point
DIAGnostic1:POINt:CATalog?
// ATT6HP_AF_DET_N,ATT6HP_AF_DET_P,ATT6HP_AF_DET_TEMP...
DIAGnostic1:MEASure:POINt? "ATT6HP_AF_DET_N"
// 0.000000V
:DIAGnostic<hw>:BGINfo:CATalog?
................................................................................ 333
..............................................................................................334
:DIAGnostic<hw>:POINt:CATalog?
.................................................................................. 334
:DIAGnostic<hw>[:MEASure]:POINt?
............................................................................... 335
:DIAGnostic<hw>:BGINfo:CATalog?
Queries the names of the assemblies available in the instrument.
DIAGnostic Subsystem