Remote Control Commands
R&S
®
SMA100B
416
User Manual 1178.3834.02 ─ 03
Parameters:
<Logarithmic>
float
The unit is mandatory
Range:
0.01 to 100
Increment: 0.01
*RST:
1
Default unit: PCT
Example:
See
Manual operation:
See
" Step Linear/Step Logarithmic "
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear]
<Linear>
Sets the step width for the linear sweep.
For information on how the value is calculated and the interdependency with other
parameters, see
Chapter 5.2.1, "Correlating Parameters in Sweep Mode"
Parameters:
<Linear>
float
Range:
0 to STOP-STARt
Increment: 0.1
*RST:
1000
Example:
See
Manual operation:
See
" Step Linear/Step Logarithmic "
12.15.6 SOURce:LIST Subsystem
The
SOURce:LIST
subsystem contains all commands for defining lists and for han-
dling of list files.
List files have the predefined file extension
*.lsw
.
Refer to
Chapter 12.5.2, "Accessing Files in the Default or in a Specified Directory"
on page 314 for general information on file handling in the default and in a specific
directory.
●
*RST
does not affect data lists.
●
SCPI refers to the individual lists as segments.
Example: Create an RF list and activate the list mode
The following example shows a command sequence to create an RF list and to acti-
vate the list mode. Further hardware settings are not considered.
// Reset the instrument to start from an initial state
// Query the available list files in the default
// directory /var/user
SOURce Subsystem