Remote Control Commands
R&S
®
SMA100B
356
User Manual 1178.3834.02 ─ 03
:READ<ch>[:POWer]?
Triggers power measurement and displays the results. The sensor returns the result in
the unit set with command
Certain power sensors, such as the R&S NRP-Z81, return two values, first the value of
the average level and - separated by a comma - the peak value.
Note:
This command does not affect the local state, i.e. you can get results with local
state on or off. For long measurement times, we recommend that you use an SRQ for
command synchronization (MAV bit).
Suffix:
<ch>
.
1..3
Return values:
<Power>
float or float,float
Example:
SENS1:UNIT DBM
Selects unit dBm for presentation of measurement result.
READ1?
Queries the measurement result of the sensor.
Response: -45.6246576745440230
-45.6 dBm were measured at the given frequency.
Example:
R&S NRP-Z81
READ1?
Response:
-55.62403263352178,-22.419472478812476
-55.6 dBm is the measured average level, -22. 4 dBm is the
measured peak level at the given frequency.
Usage:
Query only
Manual operation:
See
" Level (Peak) / Level (Average) "
:SENSe<ch>:UNIT[:POWer]
<Power>
Selects the unit (Watt, dBm or dB
μ
V) of measurement result display, queried with
.
Parameters:
<Power>
DBM | DBUV | WATT
*RST:
DBM
Example:
SENS2:UNIT DBM
Selects dBm as unit for the measured value returned by com-
mand
READ
.
READ2?
Response: 7.34
7.34 dBm are measured by sensor 2.
Manual operation:
See
" Level (Peak) / Level (Average) "
SENSe, READ, INITiate and SLISt Subsystems