Remote Control Commands
R&S
®
SMA100B
401
User Manual 1178.3834.02 ─ 03
Parameters:
<Start>
float
Range:
300kHz to RFmax
Increment: 0.01Hz
*RST:
100 MHz
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
" Start Frequency/Stop Frequency "
[:SOURce<hw>]:FREQuency:STOP
<Stop>
Sets the stop frequency range for the RF sweep.
See
Chapter 5.2.1, "Correlating Parameters in Sweep Mode"
Parameters:
<Stop>
float
Range:
300kHz to RFmax
Increment: 0.01Hz
*RST:
500 MHz
Default unit: Hz
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
" Start Frequency/Stop Frequency "
[:SOURce<hw>]:FREQuency:STEP:MODE
<Mode>
Defines the type of step size to vary the RF frequency at discrete steps with the com-
mands
Parameters:
<Mode>
DECimal | USER
DECimal
Increases or decreases the level in steps of ten.
USER
Increases or decreases the level in increments, set with the
command
.
*RST:
DECimal
Example:
// increasing the RF frequency with a step size of 50 KHz
SOURce1:FREQuency:STEP 50E3
SOURce1:FREQuency:STEP:MODE USER
SOURce1:FREQuency:CW UP
Manual operation:
See
SOURce Subsystem