Remote Control Commands
R&S
®
SMA100B
385
User Manual 1178.3834.02 ─ 03
Return values:
<Points>
integer
Range:
0 to INT_MAX
*RST:
0
Example:
See
Example "Generating a pulse train signal"
Usage:
Query only
[:SOURce<hw>]:PULM:TRAin:REPetition
<Repetition>
Sets the number of repetitions for each pulse on/off time value pair.
Parameters:
<Repetition>
Repetition#1{, Repetition#2, ...}
0 = ignore value pair
Set "Repetition = 0" to skip a particular pulse without deleting
the pulse on/off time value pair
Range:
0 to 65535
Example:
See
Example "Generating a pulse train signal"
Manual operation:
See
[:SOURce<hw>]:PULM:TRAin:SELect
<Filename>
Selects or creates a data list in pulse train mode.
If the list with the selected name does not exist, a new list is created.
Parameters:
<Filename>
string
Filename or complete file path; file extension can be omitted.
Example:
See
Example "Generating a pulse train signal"
Manual operation:
See
Pulse Train Data Exchange
Option: R&S
SMAB-K27
With the following commands, you can configure pulse trains in ASCII format and
export or import them accordingly.
SOURce1:PULM:SOURce INT
SOURce1:PULM:MODE PTR
SOURce1:PULM:TRAin:SELect "/var/user/pt"
SOURce1:PULM:STATe 1
SOURce1:PULM:TRAin:DEXChange:MODE EXP
SOURce1:PULM:TRAin:DEXChange:AFILe:EXTension TXT
SOURce1:PULM:TRAin:DEXChange:AFILe:SEParator:DECimal DOT
SOURce Subsystem