Remote Control Commands
R&S
®
SMA100B
324
User Manual 1178.3834.02 ─ 03
:CALibration:ROSCillator[:DATA]
<Data>
Sets a user-defined calibration value for the internal reference frequency.
Parameters:
<Data>
integer
Range:
0 to INT_MAX
*RST:
0
Example:
See
[:SOURce]:ROSCillator[:INTernal]:ADJust[:
:CALibration<hw>:CONTinueonerror
<State>
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
Suffix:
<hw>
.
[1]
Optional suffix
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
n.a. (factory preset: 0)
Example:
CAL:CONT ON
// Continues calibration after an error
Manual operation:
See
"Continue Adjustment on Error"
:CALibration:CSYNthesis[:MEASure]?
Starts all adjustments which affect the clock synthesis.
Return values:
<Measure>
0 | 1 | OFF | ON
Example:
CALibration:CSYNthesis:MEASure?
// starts adjustment
// 0
// Adjustment successful
Usage:
Query only
Options:
R&S
SMAB-B29
:CALibration:LEVel:BWIDth
<Bandwidth>
Parameters:
<Bandwidth>
LOW | HIGH | AUTO
*RST:
AUTO
CALibration Subsystem