Remote Control Commands
R&S
®
SMA100B
463
User Manual 1178.3834.02 ─ 03
Setting parameters:
<Key>
integer
The respective functions are disabled when the protection level
is activated. No password is required for activation of a level.
A password must be entered to deactivate the protection level.
The default password for the first level is 123456.
This protection level is required to unlock internal adjustments
for example.
Example:
// to activate protection level
SYSTem:PROTect1:STATe 1
// internal adjustments or hostname cannot be changed
// to unlock protection level 1
SYSTem:PROTect1:STATe 0,123456
// internal adjustments are accessible
Manual operation:
See
:SYSTem:SECurity:VOLMode[:STATe]
<SecPassWord>, <MmemProtState>
Activates volatile mode, so that no user data can be written to the internal memory per-
manently.
To enable volatile mode, reboot the instrument. Otherwise the change has no effect.
Parameters:
<MmemProtState>
0 | 1 | OFF | ON
*RST:
0
Setting parameters:
<SecPassWord>
string
Current security password
The default password is 123456.
Example:
SYSTem:SECurity:VOLMode:STATe "123456", 1
SYSTem:REBoot
Manual operation:
See
:SYSTem:COMMunicate:GPIB:LTERminator
<LTerminator>
Sets the terminator recognition for remote control via GPIB interface.
SYSTem Subsystem