Remote Control Commands
R&S
®
SMA100B
431
User Manual 1178.3834.02 ─ 03
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:NOISe:LEVel[:ABSolute]?
Queries the level of the noise signal in the system bandwidth within the enabled band-
width limitation.
Return values:
<Absolute>
float
Noise level within the bandwidth limitation
*RST:
3.84 MHz
Example:
See
Example "Configuring the noise generator"
Usage:
Query only
Manual operation:
See
12.15.8 SOURce:PGEN Subsystem
The
PGEN
subsystem contains the commands for setting output of the pulse modula-
tion signal.
Example: Using pulse generator as source for pulse modulation
// select pulse generator as source for pulse modulation
// enable pulse modulation
SOURce1:PULM:SOURce INT
SOURce1:PULM:STATe 1
// pulse generator and signal output are also activated
SOURce1:PGENerator:STATe?
// 1
SOURce1:PGENerator:OUTPut:STATe?
// 1
PGENerator:OUTPut:POLarity NORMal
// to disable pulse generator
SOURce1:PGENerator:STATe 0
// activate the pulse modulation of the RF carrier
SOURce1:PULM:STATe 1
[:SOURce<hw>]:PGENerator:OUTPut:POLarity
................................................................ 431
[:SOURce<hw>]:PGENerator:OUTPut[:STATe]
.................................................................432
[:SOURce<hw>]:PGENerator:STATe
............................................................................... 432
[:SOURce<hw>]:PGENerator:OUTPut:POLarity
<Polarity>
Sets the polarity of the pulse output signal.