Remote Control Commands
R&S
®
SMA100B
325
User Manual 1178.3834.02 ─ 03
:CALibration:LFOutput[:MEASure]?
Performs all adjustments which affect the internal modulation generator.
Return values:
<Measure>
0 | 1 | OFF | ON
Example:
CAL:LFO?
// 0
// the adjustments have been performed successfully
Usage:
Query only
:CALibration<hw>:ALL:DATE?
Queries the date of the most recently executed full adjustment.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Date>
string
Example:
CAL:ALL:DATE?
// "2016-01-01"
Usage:
Query only
Manual operation:
See
:CALibration<hw>:ALL:TEMP?
Queries the temperature deviation compared to the calibration temperature.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Temperature>
string
Example:
CALibration:ALL:TEMP?
// "+39.00 K"
Usage:
Query only
Manual operation:
See
"Temperature Offset To Last Full Adjustment"
:CALibration<hw>:DEBug
<State>
Activates logging of the internal adjustments.
CALibration Subsystem