Example: Configuring the reference oscillator
// Using 100 MHz external reference source
SOURce:ROSCillator:PRESet
SOURce:ROSCillator:SOURce EXT
SOURce:ROSCillator:EXTernal:RFOFf:STATe 1
SOURce:ROSCillator:EXTernal:FREQuency 100MHZ
SOURce:ROSCillator:EXTernal:SBANdwidth WIDE
// 100 MHz (loop thought) and 1 GHz output
SOURce:ROSCillator:OUTPut:FREQuency:MODE LOOP
SOURce:ROSCillator:OUTPut:ALTernate:FREQuency:MODE DER1G
SOURce:ROSCillator:INTernal:ADJust:STATe 0
// Variable external reference frequency
// SOURce:ROSCillator:EXTernal:FREQuency VAR
// SOURce:ROSCillator:EXTernal:FREQuency:VARiable 100000000
// Using the internal reference frequency
SOURce:ROSCillator:SOURce INT
SOURce:ROSCillator:INTernal:TUNing:STATe 1
SOURce:ROSCillator:INTernal:TUNing:SLOPe LOW
// 10 MHz and 1 GHz output
SOURce:ROSCillator:OUTPut:FREQuency:MODE DER10M
SOURce:ROSCillator:OUTPut:ALTernate:FREQuency:MODE DER1G
...................................................................................... 443
..................................................................................... 443
[:SOURce]:ROSCillator:INTernal:TUNing[:STATe]
.............................................................444
[:SOURce]:ROSCillator:INTernal:TUNing:SLOPe
.............................................................. 444
[:SOURce]:ROSCillator:EXTernal:RFOFf[:STATe]
............................................................. 444
[:SOURce]:ROSCillator:EXTernal:FREQuency
..................................................................444
[:SOURce]:ROSCillator:EXTernal:FREQuency:VARiable
................................................... 445
[:SOURce]:ROSCillator:EXTernal:SBANdwidth
................................................................. 445
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE
......................................................... 445
[:SOURce]:ROSCillator:OUTPut:ALTernate:FREQuency:MODE
......................................... 446
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue
..............................................................446
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe]
............................................................447
[:SOURce]:ROSCillator:PRESet
Resets the reference oscillator settings.
Example:
See
Example "Configuring the reference oscillator"
Usage:
Event
Manual operation:
See
[:SOURce]:ROSCillator:SOURce
<Source>
Selects between internal or external reference frequency.