Section 5: Pulse measure and pulse generator units
Model 4200A-SCS Parameter Analyzer Reference Manual
5-76
4200A-901-01 Rev. C / February 2017
Example: pulse_meas_wfm
This function sets channel 1 of the PMU for the waveform discrete measure type to acquire the
voltage/current readings for the waveform and the time stamps. It also enables LLEC.
pulse_meas_wfm(PMU1, 1, 0, 1, 1, 1, 1);
Where:
•
Instr_id
= PMU1
•
chan
= 1 (channel 1)
•
AcquireType
= 0 (discrete)
•
AcquireMeasV
= 1 (enable)
•
AcquireMeasI
= 1 (enable)
•
AcquireTimeStamp
= 1 (enable)
•
LLEComp
= 1 (enable)
Example: pulse_meas_timing
This function sets the following pulse measure timing settings for five spot mean measurements for
channel 1 of PMU1:
pulse_meas_timing(PMU1, 1, 0.6, 0.8, 5);
Where:
•
Instr_id
= PMU1
•
chan
= 1 (channel 1)
•
StartPercent
= 0.6 (60%)
•
StopPercent
= 0.8 (80%)
•
NumPulses
= 5 (output one pulse)