Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
818
User Manual 1178.8520.02 ─ 01
[SENSe:]ESPectrum<sb>:MSR:GSM:CARRier
<Power>
Defines the power of the GSM carrier (if available, see
This command is only available for band category 2 (see
Suffix:
<sb>
.
1..n
Parameters:
<Power>
Range:
0 dBm to 100 dBm
*RST:
39.0 dBm
Default unit: dBm
Example:
//Select the band category 2
ESP2:MSR:BCAT BC2
//Set the base station class to medium range
ESP2:MSR:CLAS MED
//Set the maximum output power to 10 dBm.
ESP2:MSR:MPOW 10
//Set the base station RF bandwidth to 20 MHz
ESP2:MSR:RFBW 20MHZ
//GSM/Edge present
ESP2:MSR:GSM:CPR ON
//Power of the GSM carrier is 20dBm
ESP2:MSR:GSM:CARR 20
//Calculate limits for MSR SEM
ESP2:MSR:APPL
Manual operation:
See
[SENSe:]ESPectrum<sb>:MSR:GSM:CPResent
<State>
This command defines whether a GSM/Edge carrier is located at the edge of the speci-
fied RF bandwidth. In this case, the specification demands specific limits for the SEM
ranges.
This command is only available for band category 2 (see
Suffix:
<sb>
.
1..n
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Configuring and Performing Measurements