Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
979
User Manual 1178.8520.02 ─ 01
MMEMory:LOAD<n>:TRACe
<Trace>, <FileName>
This command imports trace data from the specified window to an ASCII file.
Suffix:
<n>
.
Parameters:
<Trace>
Number of the trace to be stored
(This parameter is ignored for
ALL
).
<FileName>
String containing the path and name of the import file.
MMEMory:STORe<n>:TRACe
<Trace>, <FileName>
This command exports trace data from the specified window to an ASCII file.
For details on the file format see
Chapter 9.6.6, "Reference: ASCII File Export Format"
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see
Chapter 5.1.7, "Protecting Data Using the Secure User Mode"
Suffix:
<n>
.
Parameters:
<Trace>
Number of the trace to be stored
(This parameter is ignored if the option "Export all Traces and all
Table Results" is activated in the Export configuration settings,
see
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR1:TRAC 1,'C:\TEST.ASC'
Stores trace 1 from window 1 in the file
TEST.ASC
.
Manual operation:
See
" Export Trace to ASCII File "
14.8.2.7
Programming Example: Configuring a Spectrogram
This example demonstrates how to configure a spectrogram for a basic frequency
sweep in a remote environment. The spectrogram is displayed in addition to the spec-
trum display, in a new window. In addition, the usage of special spectrogram markers is
demonstrated (see
Chapter 14.8.3.6, "Marker Search (Spectrograms)"
Analyzing Measurements (Basics)